PARAMETER SETTINGS AND MAINTENANCE: Parameter Settings
●
FUNCTION (Various functions)
FUNCTION parameter screen
Parameter option
Setting contents
71. WINDOW CHECK
(Measuring window check)
YES / DAY / NO
Sets whether to automatically check the measuring window for cleanliness.
The measuring window is automatically checked for cleanliness before
measurement then the result (“OK” or “NG”) is displayed,
•
YES: The measuring window is checked every startup.
•
DAY: The measuring window is checked at the first startup of the day.
•
NO: The measuring window is not checked.
72. AUTO OFF
(Auto off)
5MIN / 10MIN / 15MIN / NO
Sets the auto off function.
The function to automatically turn off the device when the device is left idle
for the specified time.
Selectable from among “5MIN”, “10MIN”, “15MIN”, and “NO”.
Before the device is automatically turned off, the screen changes as fol-
lows:
•
After 1 minute: Returns to the main screen (the screen displayed at
startup).
•
After 4 minutes: The LCD backlight and fixation lamp are turned off (sleep mode).
* The memory indicator blinks.
•
1 minute before the set time: If data remains in the device that has not
been printed or output, five beeps sound as a warning.
•
After the set time: The device turns off.
When the device is in sleep mode, pressing any button displays the main
screen.
* When “NO” is selected, the screen returns to the main screen after 1 minute.
73. BEEP
(Beep sound)
LOW / HIGH / NO
Selects the pitch of the beep sound (electronic sound) produced during
measurement or such.
74. LCD BRIGHTNESS
(Brightness of LCD)
1 - 4 - 8
Sets the brightness of the LCD.
75. AUTO PS
(Automatic PS measurement)
YES / NO
Sets whether to measure the pupil size (PS) automatically for AR measurement.