QS
set microstep factor
IMM
PGM
MIP
USAGE
♦
♦
SYNTAX
xx
QS
nn or xx
QS
?
PARAMETERS
Description
xx
[ int ] -
axis number
nn
[ int ] -
microstep value
Range
xx
-
1
to
MAX AXES
nn
-
1
to
250
for step motors
1
to
1000
for commutated step motors
or
?
to read current setting
Units
xx
-
none
nn
-
none
Defaults
xx
missing:
error 37, AXIS NUMBER MISSING
out of range:
error 9, AXIS NUMBER OUT OF RANGE
nn
missing:
error 38, COMMAND PARAMETER MISSING
out of range:
error xx2, PARAMETER OUT OF RANGE
DESCRIPTION
This command is used to set the microstep factor for a Newport Unidrive
compatible programmable driver with step motor axis. This command must be
followed by the
QD
update driver command to take affect.
RETURNS
If the “?” sign takes the place of
nn
value, this command reports the current
setting
REL. COMMANDS
QD
-
update driver
QI
-
set maximum motor current
QT
-
set tachometer gain
QG -
set gear constant
QV
-
set average motor voltage
EXAMPLE
2QS?
|
read microstep factor of axis # 2
100
|
controller returns a
value of 100 for axis #2
2QS 250
|
set microstep factor to 250 for axis #2
2QD
|
update programmable driver with latest settings for axis #2
SM
|
save all controller settings to non-volatile memory
Section 3 – Remote Mode
3-119
Summary of Contents for ESP300 Series
Page 1: ...ESP300 Motion Controller Driver User s Manual...
Page 4: ...ESP300 1999 EU Declaration of Conformity iv Preface...
Page 29: ...This page is intentionally left blank Section 2 Modes of Operation 1 15...
Page 30: ......
Page 230: ...4 16 Section 4 Advanced Capabilities...
Page 274: ...This page is intentionally left blank 6 6 Section 6 Servo Tuning...
Page 290: ...This page is intentionally left blank B 6 Appendix B Trouble Shooting and Maintenance...