Wear protective gloves to protect yourself in case of accidental breakage.
2.
Lightly coat the pH probe with glycerol.
CAUTION!
Do not install the pH port adaptor in the headplate before inserting the
probe. Follow the steps below to fit the pH port adapter onto the probe
first, then insert the probe and adapter into the headplate.
Figure 11: pH Probe with Port Adapter (exploded)
With reference to Figure 11 above:
3.
Gently slide the top portion of the knurled port adapter (part of the probe
kit) onto the probe.
4.
Slide the two white ferrules onto the probe, the narrower one on top of the
deeper, cup-shaped one.
5.
Gently slide the bottom portion of the port adapter onto the probe, taking
care to orient the longer threaded section toward the top of the probe.
6.
Remove the two O-rings installed in the pH port; first slide the white
Teflon O-ring onto the probe, then follow with the black 12mm port
adapter O-ring.
7.
Do not yet close up all the elements of the port adapter.
CAP
pH PROBE ADAPTER
(top portion)
TOP FERRULE
BOTTOM FERRULE
pH PROBE ADAPTER
(bottom portion)
TEFLON O-RING (white)
PORT O-RING (black)