User’s Guide
1.
In the
System Settings
screen, press the
Refresh
button to update the
current software status and to search for a new USB drive.
2.
The name of the new drive folder appears in the
Update File
box.
3.
Press the
Update
button to install the file. The file will reboot twice; this
may take a little time.
4.
The Software pane will reflect the changes.
Updating software will not affect any previous user settings.
12.3 Hardware
Setup
The BioFlo/CelliGen 115 system you purchase is preset in the factory as “Unit1”
with all the accompanying hardware. In the Unit1 hardware list shown in the sample
Hardware Setup
screen (
Figure 44
), the system has the Base Power module, the
Main pH/DO module, and one TMFC. This system is also set to NBS Modbus
communication mode (
see the SCADA pane in Figure 44
), and has the Unit ID number
of 2. This is the unit’s multidrop identification number. Remember, when you add
units, that no two nodes on the network can have the same multidrop identification
number.
Figure 44: Hardware Setup Screen
To add new hardware (such as a new utility station), after you connect the module(s)
to the system:
1.
Press the
Scan Hardware
button in this screen. All new hardware scanned will
appear in the
New Hardware
box (
see Figure 45a
).
2.
Press the
>>>
button for the Unit name you wish to assign (Unit2, for example),
and the new hardware list will move into that unit’s Module box (
see Figure 45b
).
The dark blue Unit
Name tab indicates
that the hardware
shown belongs to
Unit 1.
This unit’s
multidrop number