97
New Brunswick Scientific
User’s Guide
12.2 System
Settings
Press the third tab in the
SETUP
screen to open the
System Settings
screen (
see
Figure 43
). Use this feature select the onscreen language you prefer, to reset the date
and/or time, to update the software, and to calibrate the BioFlo/CelliGen 115
touchscreen.
Figure 43: System Settings Screen
12.2.1 Resetting
Date/Time
To reset the onscreen date and/or time (displayed in the lower righthand corner
of every screen):
1.
In the System Settings screen press the edit box for the numeric parameter
you wish to change.
2.
Use the pop-up touchpad to input the new number and press the
OK
button.
3.
To change the month, press the down arrow (
T
) and press the month you
wish to select from its associated drop-down menu.
4.
Press the
Set
button to save the new information. You can do this after
each change, or after all changes have been made.
12.2.2 Updating
Software
To update the system software, obtain a new version of the software in a USB
drive and plug the drive into the USB port on the control cabinet:
Other languages
are not available
at this time.
To change
the Date
and/or Time,
see Section
16.3.1.
You will find listed here the current User
Interface and Control Program versions. To
update the software, see Section 16.3.2.
To recalibrate
the system’s
touchscreen,
press the Calib.
button, then
touch the
onscreen target
each time it
appears. You
will be guided
through the
process.