If any problems occur with your BioFlo/CelliGen 115 system or its individual components, do
not attempt to perform any service on it. Unauthorized servicing may void the warranty.
Please contact your local NBS Service Department or your local NBS distributor.
In any correspondence with NBS, please refer to the Model Number (BioFlo/CelliGen 115),
and the Manufacturing Part Number and Serial Number of the unit.
17.1 Troubleshooting
WARNING!
Always turn your BioFlo/CelliGen 115 off and disconnect the power cord
before performing maintenance.
As with any equipment, difficulties sometimes arise. If you experience a problem with
the operation of your BioFlo/CelliGen 115, consult the following list of symptoms.
You may be able to resolve the situation easily and quickly yourself.
If the problem is not listed below, or if the suggested solutions do not work, please call
your NBS representative to request a service technician.
Other than the solutions
proposed below, do not attempt to fix the equipment yourself.
Problem
Possible Solution
TEMPERATURE:
Readout is a negative value
(typically –225° C).
•
Inspect the temperature probe for obvious damage;
replace it if necessary.
•
Make sure the temperature probe is connected to the
cabinet jack.
The unit will not heat up.
•
Make sure the unit was primed at start-up.
•
Make sure the temperature probe is plugged into the
vessel thermowell.
•
Water pressure may be too low; raise pressure within
recommended range.
•
Verify correct connection (click to lock) of the water
inlet and outlet lines on the vessel heat exchanger.
•
Hit reset button on hot plate (if appropriate).
The unit is leaking water.
•
Inlet water pressure may be too high; lower pressure
within the recommended range.
•
Check for any loose connection of inlet hoses; tighten
if necessary.
...continued...