5 Service and Maintenance Procedures
4 Open top cover.
Open the top cover of the instrument to provide better access and clearance to the
lamp’s mounting box. This will make it easier to remove the lamp assembly.
5 Remove flash lamp assembly.
Slide the lamp assembly horizontally out the side of the instrument and lay it on its heat
sink fins, leaving the orange electrical cord connected.
Mounting Block
Mounting Screws
(Spring-loaded)
Flash Lamp
Removing the Flash Lamp Assembly
6 Withdraw lamp from assembly.
a
Use a 3/32" Allen wrench to remove the two screws that hold down the diamond
shaped retainer. Be careful not to lose screws.
Flashlamp Retainer
Flashlamp
Window
Copper Coil Spring
Multi-pin Socket
Withdraw Lamp