5.6 Replacing Fuses
4 Replace
fuse(s).
a
Once the fuse holder is out of the instrument, you can see that it contains two fuses.
b
It is possible that only one of the fuses may have blown. Molecular Devices
recommends that you replace both fuses, however, to ensure continued proper
operation. Pull both fuses out of the holder and discard them.
c
Insert new slow-blow rated fuses into the fuse holder. Either end of the fuse may be
forward.
5 Replace fuse box.
a
Insert the fuse box into the opening in the instrument, making sure that the fuses are
on the right side (toward the power receptacle).
b
Press the fuse box into place, making sure the cover snaps closed.
6 Reconnect the instrument.
a
Plug the power cord to the back into the instrument and then into the wall outlet.
b
Reconnect other cables previously disconnected.
FlexStation II Operator’s Manual – Rev. D
91
Summary of Contents for FlexStation II
Page 6: ...FlexStation II Operator s Manual Rev D vi ...
Page 92: ...4 Applications 86 FlexStation II Operator s Manual Rev D ...
Page 102: ...5 Service and Maintenance Procedures 96 FlexStation II Operator s Manual Rev D ...
Page 124: ...6 Troubleshooting Procedures 118 FlexStation II Operator s Manual Rev D ...
Page 134: ...7 Appendices 128 FlexStation II Operator s Manual Rev D ...