3 Operating Procedures
to resolve a different instrument problem. See part 6, Troubleshooting Procedures, for
instructions.
3 View Flex mode setting.
The SOFTmax PRO application defaults to Flex mode every
time you start the software. You can confirm this mode, if desired, by selecting
Instrument Setup from the Control drop-down menu.
You can also click on the Setup button on the Plate Section tool bar to view the
Instrument Settings screen.
Instrument Settings Screen
Notice that the [FLEX] button on the right is highlighted.
Note:
You can select a different read mode from the Instrument Settings screen by
clicking on one of the other four buttons at the top of the window. The rest of these
instructions assume you are remaining in Flex mode.
52
FlexStation II Operator’s Manual – Rev. D