6 Troubleshooting Procedures
Eight- or
Sixteen-Channel
Pipettor Head
Inside Top Panel
Opened
Location of Pipettor Head Inside Fluidics Module
6
Slide the pipettor head part way out from under the metal mounting plate, and of the
top panel opening, supporting it from underneath.
7
Pull the white spiral cord off the coil hook behind the pipettor head. This will free the
mechanism for easier removal.
Red Knob
Alignment Pins
Pipettor Head Detail
Note:
If you want to dispense liquid from the tips, skip the next two steps and see
section 6.4.6, Expelling Undispensed Fluid from Tips (next).
8
To completely remove the pipettor head, unscrew the black connector (at the end of the
white spiral cord) from its secure position around the receptacle.
9
Pull the connector out of the receptacle and take the pipettor head the rest of the way
out of the fluidics module.
110
FlexStation II Operator’s Manual – Rev. D