3 Operating Procedures
3.5.3 Timing
Important!
The Timing setting depends on several settings which appear after it in the
list. You will need to return to Timing after completing all settings to obtain the minimum
read interval.
Timing for readings can be adjusted in this window. You can enter the total run time
and the time interval between readings.
Timing Setting Dialog Box
•
To change the default values, click in the appropriate box.
•
Type in the desired run time or interval. The acceptable run time range is from zero
to a thousand (0-1000) seconds.
When you have entered the values you desire, the total Number of Reads calculated by
the instrument is displayed automatically. Depending on what sensitivity you have
selected in the previous setting box, the Minimum Interval and the Minimum Run Time
will automatically adjust. The minimum interval is also dependent on the Compound
Transfer settings and the number of wells to be read per column.
Errors:
If you select a time out of range, an error message appears:
If you select an inappropriate interval, an error message appears at the bottom of the
window:
58
FlexStation II Operator’s Manual – Rev. D