6.5 General Error Messages
6.5 General Error Messages
Error messages can appear during operation of the instrument. Follow any instructions
present in the message. Anytime a message persists, or if you cannot return the
instrument and software to normal operation, contact Technical Support and be sure to
note which error message you observed.
Note:
See section 6.4.1, Assessing a Mechanical Problem, for error messages specific to
fluidics operation failures.
General Error 1: No assay plate
If you get this message, check to be sure there is an assay plate correctly placed in the
reading chamber drawer. Start the reading again.
If you get any one of the following messages, attempt your reading again before
continuing with troubleshooting.
General Error 2: Unexpected response, result code 105
General Error 3: Communication problem, response to late
General Error 4: Unexpected response
General Error 5: Measurement aborted
FlexStation II Operator’s Manual – Rev. D