Chapter Nine: RS-232 Communications
Command Messages
283
Read Only Status Messages
Message Description
Message Format
@ command
category
command
number
Parameter
Data
Optional
checksum
Software Version
@
7
0
1
*
?
CR
c1
c2
Channel Status
@
7
0
2
*
?
CR
c1
c2
Alarm Status
@
7
0
3
*
?
CR
c1
c2
Configuration Setup
@
7
0
4
*
?
CR
c1
c2
Rear Panel Status
@
7
0
5
*
?
CR
c1
c2
Reference Voltage
@
7
0
6
*
?
CR
c1
c2
* Any character may be entered since the parameter field has no significance in the message
Table 82: Read Only Status Messages
Response Format for the Read Only Status Messages
Message
Response
Description
Software Version
<version> CR
where <version> represents the software version
number
Channel Status
C1:<text>LF CR
C2:<text> LF CR
C3:<text> LF CR
C4:<text> LF CR
where <text> is:
ON = Sensor normal
OFF = Manual off
AOFF = Auto off
DCNT = Disconnected
OVER = Overranged
UNDR = Underranged
INIT = Initializing
ZERO = Zeroing
E1 = Channel error #
- - - - = Channel not
installed
Alarm Status
AA:<text>LF CR
AB:<text> LF CR
AC:<text> LF CR
AD:<text> LF CR
where <text> is:
ENAB = Alarm
Enabled
ENAL = In alarmed
state
ENLT = Just latched
DSBL = Alarm disabled
- - - - = Channel not
installed
Table 83: Response Format for the Read Only Status Messages
(Continued on next page)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com