General Information
Chapter Seven: Operation in Setup Mode
184
Summary of Setup Mode Codes
Function
Left Display
Center Display
Right Display
Main Display
1 Select Volt/
Pressure/ or
Current
Code number
-
HU (high voltage),
EC (emission
current), or IC (ion
current)
*Main display must
be Volt, or this is
unused
Selects Volt or
PrES (pressure)
2 Select
Pressure Units
Code number
-
-
Selects pressure
unit
3 Edit Audio
Alarms
Code number
-
-
On or Off audio
alarm
4 Adjust Alpha
Code number
-
-
Alpha value
5 Set up
RS-232 Port
Code number
Data bits and
parity
Data logger or
MKS protocol
Baud rate
7 Set up Main
Display in
Leakage Mode
Code number
-
-
Determine
Leakage Mode
main display
8 Edit Data
Logger Interval
Code number
-
-
Data logger
interval value
10x Adjust
Display Lag
Code number
-
-
Display lag
percent
11x Auto Zero
Code number
-
On/Off Auto Zero
Reference channel
number
12x Adjust Ion
Gauge Auto
Power Control
Code number
& channel
number for ion
gauge
Reference
channel number
On/Off auto power
control
Trip point for auto
power control
13x Adjust
Softstart
Code number
and recipe
number
-
On/Off softstart
speed function
Softstart speed
14x Set
Polarity &
Control
Channel
Code number
and recipe
number
-
Control channel
number recipe
dependent
Select dir (direct)
or rEV (reverse
polarity
Table 46: Summary of Setup Mode Codes
(Continued on next page)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com