Hot Cathode Messages
Appendix G: Board Specific Commands
322
Hot Cathode Messages
The 146 unit supports up to four hot cathode gauges.
Hot Cathode Messages
Function
Description
Command
Range and Restrictions
Power
Channel
power
@081<para>:<state>
<para> is the channel number 1, 2, 3, or 4
<state> is ON for on; OFF for off
Sensor must be installed
Auto power
function
@082<para>:<state>
<para> is the channel number 1, 2, 3, or 4
<state> is E for enabled; D for disabled
Auto power
reference
@083<para>:<ref>
<para> is the channel number 1, 2, 3, or 4
<ref> is the reference channel number
Auto power
shutoff
pressure
@084<para>:<pres>
<para> is the channel number 1, 2, 3, or 4
<pres> is the shutoff pressure, in ASCII
Degas
@085<para>:<dgas>
<para> is the channel number 1, 2, 3, or 4
<dgas> is:
H for high power degas
L for low power degas
O for degas function off
Configuration
Select gauge
factor
@064<para>:<fact >
<para> is the channel number 1, 2, 3, or 4
<fact> is a numeric ASCII string
Disconnect
pressure
@065<para>:<pres>
<para> is the channel number 1, 2, 3, or 4
<pres> is the disconnect pressure, in ASCII
High pressure
shutoff
@066<para>:<pres>
<para> is the channel number 1, 2, 3, or 4
<pres> is the shutoff pressure, in ASCII
Fail-Safe
function
@067<para>:<state>
<para> is the channel number 1, 2, 3, or 4
<state> is E for enabled; D for disabled
(Continued on next page)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com