How To Span a Sensor with a Reference
Chapter Four: Operation in Normal Mode
116
1. Repeatedly press the [
DISPLAY MODE
] key until the 146 unit is in Normal Mode.
The system responds by scrolling through modes.
2. Press [
SPAN
/.], followed by [5/
DUAL
], followed by [1], [2], [3], or [4] (lower range
sensor), followed by [1], [2], [3], or [4] again (upper range sensor).
The system responds by spanning the sensor on the first channel, to the reading of the
sensor on the second channel.