Chapter Five: Operation in Leakage Mode
General Information
125
Chapter Five: Operation in Leakage Mode
General Information
Press the [
DISPLAY MODE
] key once from Normal Mode to get to Leakage Mode. The word
LEAKAGE
appears in small letters above the main display.
Torr
m
LEAKAGE
HOLD
TORR
4
LEFT DISPLAY
CENTER DISPLAY
RIGHT DISPLAY
The pressure reading from
the gauge on channel 1 is
shown here.
The pressure reading from
the gauge on channel 3 is
shown here.
The pressure reading from
the gauge on channel 2 is
shown here.
MAIN DISPLAY
This is the leakage rate measured from the gauge on channel 4.
Note: The main display may show the Leakage Rate (as in this
example) or instantaneous Leakage.
Figure 33: Leakage Mode
Figure 33 shows a sample window for Leakage Mode. In Leakage Mode, all upper displays
show a pressure reading, provided the displays are hooked up to active sensors and the sensors
are reading within their range. The main display shows either leakage rate, or instantaneous
leakage.
Leakage rate is the leakage over a user-defined time base. The leakage rate display shows the
rate in pressure units/second. For example, the leakage rate may be 3 Torr per 30 seconds, where
Torr is the pressure unit, and 30 seconds is the user-defined time base. The display shows the
rate as .1 Torr because a rate of 3 Torr/30 seconds equals a rate of 0.1 Torr/second. Refer to
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com