Chapter Seven: Operation in Setup Mode
Code 7: How To Set the Display for Leakage Mode
223
1. Repeatedly press the [
DISPLAY MODE
] key until the 146 unit is in Setup Mode.
The system responds by scrolling through modes.
2. Use the arrow keys to scroll to the
CODE
legend.
The system responds by causing the left display to blink.
3. Enter the number 7 and press the [
ENTER
] key.
The system responds by displaying the appropriate parameters for code 7, and causing
the main display to blink.
4. Enter the leakage rate time base OR 0 for instantaneous leakage. The default value of 0
configures the 146 unit for instantaneous leakage display.
A. If the display is correct, press the [
ENTER
] key.
B. If the display is not correct, enter 0 for instantaneous leakage, or a number from 1 to
60 (seconds) to define the time base for leakage rate calculations, then press the
[
ENTER
] key.
The system responds by accepting the value entered.