Auxiliary Output Messages
Appendix G: Board Specific Commands
328
Auxiliary Output Messages (Continued)
Function
Description
Command
Range and Restrictions
Analog
Outputs
Channel
Output Type
@071<para>:<type>
<para> is the analog output (1, 2, or 3)
<type> is LN for linear
LO1 to LO9 for log 1 to log 9
SP for set point
Input Channel
Selection
@072<para>:<chan>
<para> is the analog output (1, 2, or 3)
<chan> is the input channel (1, 2, 3, 4, 5
(dual 1), or 6 (dual 2))
FS Pressure
(10 V output)
@073<para>:<pres>
<para> is the analog output (1, 2, or 3)
<pres> is the full scale pressure