Chapter Seven: Operation in Setup Mode
Code 11x: How To Turn Auto Zeroing ON or OFF
229
1. Repeatedly press the [
DISPLAY MODE
] key until the 146 unit is in Setup Mode.
The system responds by scrolling through modes.
2. Use the arrow keys to scroll to the
CODE
legend.
The system responds by causing the left display to blink.
3. Enter the number 11x (where x represents the channel number of the gauge to be
zeroed), and press the [
ENTER
] key.
The system responds by showing code 11x in the left display, and causing the main
display to blink.
4. Enter the reference channel number (the default is channel 1).
A. If the display is correct, press the [
ENTER
] key.
B. If the display is not correct, enter the correct reference channel number (1 through 4),
and press the [
ENTER
] key.
The system responds by accepting the reference channel number and causing the right
display to blink.
5. Select On or Off to Enable or Disable the Auto Zero function.
A. If the display is correct, press the [
ENTER
] key.
B. If the display is not correct, use the [
ON
/
LEAD
] or [
OFF
/
GAIN
] key to enable or
disable the Auto Zero function, then press the [
ENTER
] key.
The system responds by enabling or disabling the Auto Zero function for the specified
reference and selected channels.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com