System Diagnosis and Support
8-11
8.3.6
PHV Related Alarming
replace with new battery
please change the
battery
/
Battery I2C error
,
Shutdown State
Replace battery and confirm
the malfunction module.
/
Battery Hot Plug
,
Shutdown State
Stop the illegal operation.
Alarming tips
LOG record
Potential reason
Alarm!
High-voltage
transmission is
abnormal, and
images display
normally!
PHVX supply voltage error, PHVX
is XX(upper limit is XX)
Something is wrong with the programmed
voltage of power module, which make
either of the both PHV circuits or multi-path
voltage 1.1 times more than the upper limit
of voltage (100%AP
)
.
none
PHV protect, PHV1P is XX,
PHV1N is XX
PHV protection and the voltage of PHV is
less than the lower limit.
95V Power supply voltage error,
the voltage is XX ,normal region is
90V to 100V
Something is wrong with the programmed
voltage of power module: positive/negative
95V of power transmitting exceeds the
limited value
-95V Power supply voltage error,
the voltage is XX ,normal region is
-90V to -100V