9-8 Care and Maintenance
9.3.3
Peripherals and Options Check
If the system is not configured with any module or peripheral, the corresponding items checking can
be skipped.
Table 9-7 Options, Peripherals and Accessories Check list
No.
Content
Method
1.
Color and B/W video
printer
Check if the output of video printer is normal.
2.
Graph / text printer
Check if the output of graph / text printer is normal.
3.
Foot switch
Check if the foot switch can implement the set functions
according to the program.
4.
DVD-R/W
Check if DVD can work normally (write, read and pop).
5.
DICOM
Check if DICOM can work normally and send pictures and
other data to DICOM server.
Remark
:
Please refer to 5.3 for details.
9.3.4
Mechanical Safety Inspection
Mechanical safety inspection is mainly used to check mechanical strength and mechanical
function of the key assembly of ultrasonic system. The mode of test evaluation mainly is: Perform
the evaluation by means of visual check and operating check, if the check result cannot pass, the
system is in abnormal status now. Stop using the system and adopt proper measures. The test flow
is as following:
Check handle
Check control panel And
turning mechanism
Check display and turning
mechanism
Check other mechanics
Check the connection of
casters
Check casters
Fig 9-4 Mechanical Safety Inspection Flow
Table 9-8 Mechanical Safety Check
Summary of Contents for Z6
Page 2: ......
Page 14: ......
Page 16: ...2 2 Product Specifications Left side 9 4 7 5 6 8 ...
Page 72: ......
Page 119: ...Structure and Assembly Disassembly 7 35 图 7 59 Disassembly of Speaker 8 ...
Page 120: ......
Page 132: ......
Page 142: ......
Page 152: ......
Page 168: ......
Page 171: ...Phantom Usage Illustration B 3 ...
Page 172: ......
Page 173: ...P N 046 010452 00 5 0 ...