A-10 Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
6-
Patient Leakage Current
the 601PRO.
6) Press the print data key at any time to generate a printout of the latest measurement.
Figure 2 patient leakage Current
Note
1, In addition to Probes ,Patient leakage current test should be perform if ECG or PCG parts
used;
2, If the current test standard being used does not include Patient Leakage DC readings, or
the DC option is not enabled, then DC readings will not be available through the APPLIED PART
SOFT KEY selections. Refer to Chapter 8, Standards and Principles.
Failure
Check any broken of the Applied parts. Replace any defective one.
Check any broken of the ECG/PCG module if used, Replace any defective one.
Check any broken of the Power Unit. Replace a new one if any portion defective.
Inspect wiring for bad crimps, poor connections, or damage.
Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the
user or owner to correct any deviations. As a work around, check the other outlets to see if they
could be used instead.
Change another probe to confirm if the fail is caused by console.
Inspect wiring for bad crimps, poor connections, or damage.
If the leakage current measurement tests fail on a new unit and if situation cannot be
corrected, submit a Safety Failure Report to document the system problem. Remove unit from
operation.
If all else fails, stop using and inform the Customer Service Engineer for analysis and
disposal.
Summary of Contents for Z6
Page 2: ......
Page 14: ......
Page 16: ...2 2 Product Specifications Left side 9 4 7 5 6 8 ...
Page 72: ......
Page 119: ...Structure and Assembly Disassembly 7 35 图 7 59 Disassembly of Speaker 8 ...
Page 120: ......
Page 132: ......
Page 142: ......
Page 152: ......
Page 168: ......
Page 171: ...Phantom Usage Illustration B 3 ...
Page 172: ......
Page 173: ...P N 046 010452 00 5 0 ...