A-14 Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
7-
Mains on Applied Part Leakage
could be used instead.
Change another probe to confirm if the fail is caused by console.
Inspect wiring for bad crimps, poor connections, or damage.
If the leakage current measurement tests fail on a new unit and if situation cannot be
corrected, submit a Safety Failure Report to document the system problem. Remove unit from
operation.
If all else fails, stop using and inform the Customer Service Engineer for analysis and
disposal.
LIMITS
All countries:
For BF ECG input and transducer:
5000μA