10-8 Troubleshooting of Regular Malfunctions
speed in system preset;
Degenerated trackball
performance
Replace the trackball module
4
All key of control panel are
normal,
Single encoder failed.
Single encoder malfunction
Replace the encoder board
5
All key of control panel are
normal,
Single key of TGC failed.
TGC board malfunction
Replace TGC board
All keys of TGC failed.
Control panel malfunction
Replace Control panel
10.7 LCD Display
10.7.1 Module or Board Related
No.
Descriptions
Remarks
1
Monitor assembly
Failure rate: low
2
IO board
Failure rate: low
3
Main board
4
CPU module
10.7.2 Key Points Supporting Troubleshooting
No.
Key
Points
Supporting
Troubleshooting
Remarks
1
Backlight of the LCD
More obvious in dark
2
VGA output port on IO board
Need external LCD
3
Display mode of the LCD
Blank screen, or “No Signal” displayed, or
disordered screen, etc.
10.7.3 Troubleshooting
No.
Fault Description
Cause Analysis
Measure
1
Control panel powered on normally;
No display (blank screen) on the LCD;
Displays normally with external LCD
connecting by VGA.
Monitor
malfunction
Replace the monitor
assembly
2
Control panel powered on normally;
Main board or
Replace the main
Summary of Contents for Z6
Page 2: ......
Page 14: ......
Page 16: ...2 2 Product Specifications Left side 9 4 7 5 6 8 ...
Page 72: ......
Page 119: ...Structure and Assembly Disassembly 7 35 图 7 59 Disassembly of Speaker 8 ...
Page 120: ......
Page 132: ......
Page 142: ......
Page 152: ......
Page 168: ......
Page 171: ...Phantom Usage Illustration B 3 ...
Page 172: ......
Page 173: ...P N 046 010452 00 5 0 ...