8-10 System Diagnosis and Support
8.3.4
Fan Alarming
8.3.5
Battery Alarming
[VVV] represents the current value, and [LLL]
represents the first limit of CPU temperature.
Temperature
alarming, XX
Shut down
(XX means
inversion timing,
starting 60S)
The log records are the same
[VVV] represents the current value, and [LLL]
represents the second limit of CPU temperature.
The second
temperature of CPU
Suggestion as above
Alarming tips
LOG record
Suggestion
Fans need
maintenance,
please
contact
service
engineers
×××: System Monitor: Fan alert
!
[XXX], Current speed:
[VVV] degree, Limit speed: [LLL] rpm.
×××represents time, e.g. 2010-12-25 14:15:25
[XXX] represents fan name, [VVV] represents the current
value, and [LLL] represents the limit. The detailed fan
names are as following:
Main Board Fan
、
Power Fan
Replace the
fan/connect the line
again/remove the
barrier which causes
the fan malfunction
Alarming tips
LOG record
Suggestion
Alarming! Illegal operating battery!
Lead to permanent damage.
Battery Hot Plug
Stop the illegal operation.
Battery communication is abnormal.
Battery I2C error
Check the battery connection
or replace battery
Battery communication is abnormal,
please connect to AC power supply or
shut down! Otherwise the system
shuts down automatically after 60s!
Battery I2C error, the
battery supply
Check the battery connection
or replace battery
Battery temperature is too high; please
connect to AC power supply or shut
down! Otherwise the system shuts
down automatically after 60s!
Battery temperature is
out of range temp
,
Battery temperature is:
xx centigrade
Check if the heat radiating
condition is normal.
Battery temperature is too low; please
connect to AC power supply or shut
down! Otherwise the system shuts
down automatically after 60s!
battery voltage is low,
battery volt is xx V
Check the battery connection
or replace battery
Battery error! Battery cannot be used!
battery break
Replace battery
Battery voltage is too low, please
connect to AC power supply or shut
down! Otherwise the system shuts
down automatically after 60s!
Low battery power
,
battery power is XX
Input AC, check if the battery
charging is normal, and it can
be used when full of power.
Otherwise replace battery
Battery life is approaching, please
battery cycle is XX,
Replace battery
Summary of Contents for Z6
Page 2: ......
Page 14: ......
Page 16: ...2 2 Product Specifications Left side 9 4 7 5 6 8 ...
Page 72: ......
Page 119: ...Structure and Assembly Disassembly 7 35 图 7 59 Disassembly of Speaker 8 ...
Page 120: ......
Page 132: ......
Page 142: ......
Page 152: ......
Page 168: ......
Page 171: ...Phantom Usage Illustration B 3 ...
Page 172: ......
Page 173: ...P N 046 010452 00 5 0 ...