Troubleshooting Menus
Repair Information
3 - 42
3.16.4
Temperature Troubleshooting
SpO
2
: Communication Error
1
The monitor and the
SpO
2
modules are not
communicating properly.
1
Contact Technical
Support.
SpO
2
: Communication Stop
1
As the SpO
2
module
communication stops, the
data packets sent by the
SpO
2
module can not be
received.
1
Contact Technical
Support.
MESSAGE/ PROBLEM
REASON
SOLUTION
MESSAGE/ PROBLEM
REASON
SOLUTION
Temperature Probes not
Working
1
Poor contact from probes
to body.
1
Check the body surface
contact at the probe tip.
2
Reposition or apply
thermoconductive gel.
Temperature not displayed
1
Improper display setup.
2
Cable not plugged in.
1
Check display setup in
Monitor Setup Menu and
change as desired.
2
Check the cable.
Temp: Communication Stop
1
As the Temp module
communication stops, the
data packets sent by the
Temp module can not be
received.
1
Contact Technical
Support.
Temp: Communication Error
1
Temp module
communication error. The
command can not be
send correctly.
1
Contact Technical
Support.
Temp: SelfTest Error
1
Module sends out error
during selftest because of
some unspecified
reasons.
1
Contact Technical
Support.
Temp: Calibration Error
1
A calibration failed.
1
Restart the monitor.
2
Contact Technical
Support.
Temp: Initialization Error
1
During the Temperature
module power-on, as the
Temperature module
communication stops,
system fails to
communicate with
module.
1
Contact Technical
Support.