Disassembly Instructions
Repair Information
3 - 12
3.5.10
Removal of the Parameter Front Panel Assembly
1.
Remove the multi-parameter board assembly as stated in “Removal of the Multi-
parameter Board Assembly” on page 3-10.
2.
Unplug the tubing from the NIBP connector.
FIGURE 3-15
Unplug the tubing from the NIBP connector
3.
Unfasten the two screws that secure the SpO
2
connector and then remove the connector
(if Masimo or Nellcor SpO
2
module is configured).
4.
Remove the three screws that secure the parameter front panel.
5.
Remove the parameter front panel.
FIGURE 3-18
Remove the parameter front panel
FIGURE 3-16
Masimo SpO
2
FIGURE 3-17
Nellcor SpO
2