Disassembly Instructions
Repair Information
3 - 10
3.
Carefully remove the knob with a pair of pliers that have protection on the jaws. Some
pliers may damage the knob.
4.
Remove the nut that secures the encoder.
5.
Remove the encoder carefully.
FIGURE 3-12
Remove the encoder
3.5.9
Removal of the Multi-parameter Board Assembly
1.
Remove the front housing assembly as stated in “Removal of the Front Housing
2.
Disconnect the 50-pin ribbon cable from the interface board.
3.
Disconnect the cable from J15 of the interface board.
4.
Disconnect the NIBP communication cable from the interface board.
FIGURE 3-13
Disconnect the cables from the multi-parameter board