3 - 15
Repair Information
Disassembly Instructions
3.5.15
Removal of the NIBP Assembly
1.
Remove the multi-parameter board assembly as stated in “Removal of the Multi-
parameter Board Assembly” on page 3-10.
2.
Unplug the tubing from the NIBP gas nipple.
3.
Disconnect the NIBP communication cable from the NIBP board.
4.
Unfasten the screws that secure the NIBP assembly to the multi-parameter board
installation frame.
5.
Remove the NIBP assembly.
FIGURE 3-24
Remove the NIBP assembly
3.5.16
Removal of the Recorder Assembly
1.
Remove the two screws that secure the recorder.
2.
Release the two clips and pull out the recorder.
FIGURE 3-25
Remove the recorder assembly
3.
Disconnect the recorder cable.
4.
Remove the recorder.