Introduction
Notes
vi
CAUTION:
When cleaning the monitor, do not allow cleaning solutions
into the vent openings.
CAUTION:
When cleaning sensors, do not use excessive amounts of
liquid. Wipe the sensor surface with a soft cloth dampened
with cleaning solution. Do not attempt to sterilize.
CAUTION:
Some disinfectants may cause skin irritation. Please rinse
cuff thoroughly with water to remove any residual
disinfectants.
CAUTION:
Using dark colored soaks may stain the cuffs. Test a single
cuff to ensure that no damage will occur.
CAUTION:
When ironing or pressing the cuffs, be aware that the
Velcro
®
fasteners can melt at temperatures above 325°F
(162°C).
CAUTION:
Disposable cuffs can be cleaned using a mild soap solution
and dried with a clean cloth.
CAUTION:
Do not clean the Gas Module while it is on and/or plugged
in.
CAUTION:
The internal sampling system of the Gas Module does not
need to be cleaned or sterilized. There is no reverse flow
back to the patient. If the internal sampling system is
suspected to be clogged or dirty, the module should be
serviced by an authorized service person only.
CAUTION:
To avoid permanent damage, do not expose metal
components (e.g., pins, sockets, snaps) to disinfectants,
soaps, or chemicals.
CAUTION:
Recharge batteries in the Passport V.
CAUTION:
Remove the batteries if the Passport V is not likely to be
used for an extended period of time.
CAUTION:
Never pull the recorder paper with force when a recording
is in process; it may cause damage to the recorder.
CAUTION:
Do not leave the recorder door open unless reloading paper
or troubleshooting.
Notes
NOTE:
Unauthorized servicing may void the remainder of the
warranty. Check with the factory or with a local authorized
representative to determine the warranty status of a
particular instrument.