3 - 1
3.0
Repair Information
3.1
Introduction
This chapter provides the necessary technical information to perform repairs on the
Passport V
. The most important prerequisites for effective troubleshooting are a thorough
understanding of the instrument functions as well as understanding the theory of operation.
3.2
Safety Precautions
In the event the instrument covers are removed, observe the following warnings and
guidelines.
1.
Do not short component leads together.
2.
The instrument covers must not be removed by anyone other than qualified technical
personnel who have received supplementary instructions regarding maintenance of
medical equipment or have equivalent experience in this area.
WARNING: Internal Electrical Shock Hazard - This unit does not contain
any user-serviceable parts. Do not remove instrument
covers. Refer Servicing to qualified personnel.
WARNING: Whenever the monitor is opened for calibration or repair, a
risk (leakage) current safety check and a verification of
basic functions of all parameters should be performed