Function and Performance Checking Method 5-1
5
Function and Performance
Checking Method
5.1 NOTE
The chapter supplies the detailed method for product main function and performance checking.
This is used for referring or studying by engineer but not required.
5.2 System Running Status
5.2.1 Running Status
1. Normal power on/off operation (duration time is normal), no abnormal sounds or phenomena
occur during normal operation.
2. After ultrasound system is turned on, the fan starts working and no abnormal sound when the
fan is working.
3. Check if configuration, software version are normal through the [About] in preset menu.
4. Check if contrast and brightness of the monitor are normal.
5. Check if time and date are valid and correct.
6. Check if all status indicators are normal.
7. Check all log records with user, to confirm if there is any abnormality.
5.2.2 Working Condition
Check the ambient temperature and humidity to see if meet the requirement. The measurements
related to safety features are particularly sensitive to humidity. If the insulation feature of the system
deteriorates due to the increase of system service time or system malfunctions, the fluctuation
range of measurement results are likely to increase with the increase of humidity.
Summary of Contents for DC-68
Page 2: ......
Page 12: ......
Page 18: ......
Page 30: ......
Page 45: ...System Installation 3 15...
Page 54: ......
Page 72: ......
Page 90: ......
Page 108: ...8 2 Field Replaceable Unit 8 1 Explosive View A0 B0 C0 D0 E0 F0 G0...
Page 121: ...Field Replaceable Unit 8 15 8 2 5 Main Unit Assembly E0 E2 E4 E6 E5 E9 E8 E3 E7 E1...
Page 123: ...Field Replaceable Unit 8 17 E13 E22 E10 E17 E16 E14 E18 E19 E23 E20 E21 E15 E11 E12 E24...
Page 134: ......
Page 180: ......
Page 212: ......
Page 254: ......
Page 258: ......
Page 280: ......
Page 281: ...P N 046 011984 00 7 0...