Description of Self Test Items C-19
Test content
Read the value of TEE temperature signal. The collecting of the TEE angel signal works well if the
value keeps stable.
Analysis to test failure
The drive goes wrong if the test result appears
Error
.
The interconnection of TEE angel signal goes wrong if the test appears FAIL.
Suggestion to failure test
Restart the device to perform the self test if the test result appears
Error.
It is necessary to restore
the device (OS+doppler) if
Error
re-appears.
It is recommended to replace 4D&TEE board if the test result is
FAIL
.
C.1.41
Z1401
DSP FPGA and ECG Board Interconnection
Test
Top test item
PC module and DSP FPGA interconnection test
Test content
The program decides whether ECG is in the place, sends orders via ECG serial port, judges the
returned data, and parses Bootloader version, board ID and software version.
Analysis to test failure
The drive goes wrong if the test result appears
Error
.
ECG board is not in the place if the information “
ECG board is not in the place
” appears or the
circuit is not in the place.
The communication between ECG board and main board goes wrong if the information “
Failed to
read the data on ECG board
”.
Suggestion to failure test
Restart the device to perform the self test if the test result appears
Error.
It is necessary to restore
the device (OS+doppler) if
Error
re-appears.
It is recommended to replace ECG board if the test result is
FAIL
.
C.1.42
Z1402
ECG Board Self Test
Top test item
Communication test between DSP FPGA and ECG board.
Test content
ECG sends self test order via ECG serial port when it is in the place and analyzes the self test
result, and decides the status for A/D, Flash, DRAM, CPU, Watchdog, Voltage 33.
Analysis to test failure
The drive goes wrong if the test result appears
Error
.
ECG board goes wrong if the test result is
FAIL
.
Suggestion to failure test
Restart the device to perform the self test if the test result appears
Error.
It is necessary to restore
the device (OS+doppler) if
Error
re-appears.
It is recommended to replace ECG board if the test result is
FAIL
.
Summary of Contents for DC-68
Page 2: ......
Page 12: ......
Page 18: ......
Page 30: ......
Page 45: ...System Installation 3 15...
Page 54: ......
Page 72: ......
Page 90: ......
Page 108: ...8 2 Field Replaceable Unit 8 1 Explosive View A0 B0 C0 D0 E0 F0 G0...
Page 121: ...Field Replaceable Unit 8 15 8 2 5 Main Unit Assembly E0 E2 E4 E6 E5 E9 E8 E3 E7 E1...
Page 123: ...Field Replaceable Unit 8 17 E13 E22 E10 E17 E16 E14 E18 E19 E23 E20 E21 E15 E11 E12 E24...
Page 134: ......
Page 180: ......
Page 212: ......
Page 254: ......
Page 258: ......
Page 280: ......
Page 281: ...P N 046 011984 00 7 0...