13-12 Troubleshooting of Regular Malfunctions
13.8 ECG Module Failure
13.8.1 Related Modules or Boards
No.
Description
Remarks
1
PCBA on ECG board
/
2
PCBA on engine board
/
13.8.2 Key Points Supporting Troubleshooting
No.
Key Points Supporting Troubleshooting
Remarks
1
Wave features of ECG signal
/
2
Self test for ECG Module Information Read Test.
Confirm whether the signal
processing board communicates
normally with ECG module.
3
Self test for ECG Module Self Test.
Self test for ECG module.
13.8.3 Troubleshooting for ECG Module
No. Fault
Description
Cause Analysis
Solution
1
Fail to open
ECG relevant
functions.
ECG_12 V provides power for ECG
module. If ECG is disabled, it indicates
communication error of ECG serial port,
or connection error of ECG module. The
error of ECG module or circuit error of
ECG serial port for signal processing
board may also occur. The possibility of
the error for ECG module is
comparatively higher.
Self test for the serial
communication between
engine board and ECG
module;
Check ECG connection cable
or replace ECG assembly.
2
Enable ECG.
The waveform
of ECG appears
as a signal line
or the shape of
the wave is
abnormal.
Abnormal ECG lead or ECG module
Self test for ECG module;
Replace ECG lead or ECG
module.
Summary of Contents for DC-68
Page 2: ......
Page 12: ......
Page 18: ......
Page 30: ......
Page 45: ...System Installation 3 15...
Page 54: ......
Page 72: ......
Page 90: ......
Page 108: ...8 2 Field Replaceable Unit 8 1 Explosive View A0 B0 C0 D0 E0 F0 G0...
Page 121: ...Field Replaceable Unit 8 15 8 2 5 Main Unit Assembly E0 E2 E4 E6 E5 E9 E8 E3 E7 E1...
Page 123: ...Field Replaceable Unit 8 17 E13 E22 E10 E17 E16 E14 E18 E19 E23 E20 E21 E15 E11 E12 E24...
Page 134: ......
Page 180: ......
Page 212: ......
Page 254: ......
Page 258: ......
Page 280: ......
Page 281: ...P N 046 011984 00 7 0...