Electrical Safety Inspection A-13
ELECTRICAL SAFETY INSPECTION
7-
Mains on Applied Part Leakage
Performance
1) From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601
2) Attach the applied parts to the 601PRO applied part terminals.
3) Attach the red terminal lead to a conductive part on the DUT enclosure.
4) Press shortcut key 7. The Mains on Applied Part test is displayed.
5) Select the desired outlet configuration and applied part to test using the appropriate SOFT
KEYS:
6) Press START TEST (SOFT KEY 1) to begin the test.
7) Press the print data key to generate a printout of the latest measurement.
Figure 3 Mains on Applied part leakage
NOTE
1, In addition to Probes ,Patient leakage current test should be perform if ECG or PCG parts
used;
2, If all of the applied parts correspond to the instrument type, the applied parts will be tied
together and one reading will be taken. If any of the applied parts differ from the instrument type, all
applied parts will be tested individually, based on the type of applied part. This applies to Auto and
Step modes only.
Failure
Check any broken of the Applied part. Replace any defective one.
Check any broken of the ECG/PCG module if used, Replace any defective one.
Summary of Contents for DC-68
Page 2: ......
Page 12: ......
Page 18: ......
Page 30: ......
Page 45: ...System Installation 3 15...
Page 54: ......
Page 72: ......
Page 90: ......
Page 108: ...8 2 Field Replaceable Unit 8 1 Explosive View A0 B0 C0 D0 E0 F0 G0...
Page 121: ...Field Replaceable Unit 8 15 8 2 5 Main Unit Assembly E0 E2 E4 E6 E5 E9 E8 E3 E7 E1...
Page 123: ...Field Replaceable Unit 8 17 E13 E22 E10 E17 E16 E14 E18 E19 E23 E20 E21 E15 E11 E12 E24...
Page 134: ......
Page 180: ......
Page 212: ......
Page 254: ......
Page 258: ......
Page 280: ......
Page 281: ...P N 046 011984 00 7 0...