Troubleshooting of Regular Malfunctions 13-5
13.3.3 Image Troubleshooting
No.
Fault
Description
Cause Analysis
Solution
1
No echo in
ultrasound
image. The
probe can
recognize but
provides no
echo.
PHV voltage output of power module is 0
V or abnormal;
Errors on TR64 board, DC-DC board or
probe board;
The possibility of error on
TR64 board is higher than
DC-DC board’s. The
possibility of error on the
probe board is lower.
Confirm the error by
replacing TR64 board,
DC-DC board or probe
board.
2
Dark strips
appear on B
image
Probe malfunction, e.g., array corruption,
etc.
Replace the probe to eliminate the error.
Replace the probe;
Observe the image. The dark strips are
fixed for different probes and sockets.
Replace TR64 board;
If the strip is not fixed well for a single
probe, the probe board may go wrong.
Replace probe assembly
3
Noise appears
in B image
Probe malfunction. Replace with different
probes to check the error.
Replace the probe;
Other electrical equipment in the voltage
is working, so that the probe may be
influenced by common-mode interference
on the floor;
Ripple-shaped interference signal
appears on the image
Confirm the cause of the
failure by turning off
electrical equipment of all
peripherals connected to the
system.
Summary of Contents for DC-68
Page 2: ......
Page 12: ......
Page 18: ......
Page 30: ......
Page 45: ...System Installation 3 15...
Page 54: ......
Page 72: ......
Page 90: ......
Page 108: ...8 2 Field Replaceable Unit 8 1 Explosive View A0 B0 C0 D0 E0 F0 G0...
Page 121: ...Field Replaceable Unit 8 15 8 2 5 Main Unit Assembly E0 E2 E4 E6 E5 E9 E8 E3 E7 E1...
Page 123: ...Field Replaceable Unit 8 17 E13 E22 E10 E17 E16 E14 E18 E19 E23 E20 E21 E15 E11 E12 E24...
Page 134: ......
Page 180: ......
Page 212: ......
Page 254: ......
Page 258: ......
Page 280: ......
Page 281: ...P N 046 011984 00 7 0...