C-12 Description of Self Test Items
Suggestion to failure test
Replace the control panel if LED brightness does not respond.
C.1.27 Z0901 Primary LCD Monitor I2C Interconnection
Test
Top test item
Communication self test between PC Module and MF FPGA
Test content
Test whether I2C communication serial port between MF FPGA and primary LCD monitor works
well. The test generates primary LCD monitor’s saver, and then restores it. Primary LCD monitor’s
brightness decreases, and then returns back. Read the hardware and software versions of
primary LCD monitor.
Analysis to test failure
The drive goes wrong if the test result appears
Error
.
The interconnection between DSP FPGA and primary LCD monitor goes wrong if the test result is
FAIL
.
Suggestion to failure test
Restart the device to perform the self test if the test result appears
Error.
It is necessary to restore
the device (OS+doppler) if
Error
re-appears.
Check whether the connecting material between primary LCD monitor and main unit is fixed well,
and then replace primary LCD monitor if the test result is
FAIL
.
C.1.28
Z0402
Primary LCD Monitor Serial Number Test
Top test item
Primary LCD Monitor I2C Interconnection Test
Summary of Contents for DC-68
Page 2: ......
Page 12: ......
Page 18: ......
Page 30: ......
Page 45: ...System Installation 3 15...
Page 54: ......
Page 72: ......
Page 90: ......
Page 108: ...8 2 Field Replaceable Unit 8 1 Explosive View A0 B0 C0 D0 E0 F0 G0...
Page 121: ...Field Replaceable Unit 8 15 8 2 5 Main Unit Assembly E0 E2 E4 E6 E5 E9 E8 E3 E7 E1...
Page 123: ...Field Replaceable Unit 8 17 E13 E22 E10 E17 E16 E14 E18 E19 E23 E20 E21 E15 E11 E12 E24...
Page 134: ......
Page 180: ......
Page 212: ......
Page 254: ......
Page 258: ......
Page 280: ......
Page 281: ...P N 046 011984 00 7 0...