Electrical Safety Inspection A-7
ELECTRICAL SAFETY INSPECTION
5-
Earth Leakage Test
OVERVIEW
Run an Earth Leakage test on the device being tested before performing any other leakage
tests.
Leakage current is measured the following ways:
♦
Earth Leakage Current, leakage current measured through DUT outlet Earth
♦
Earth Leakage Current AP-EARTH (ALL Applied Parts connected to Earth), leakage current
measured through DUT outlet Earth
There is no need to attach a test lead; the 601PRO automatically connects the measuring device
internally.
TEST PROCEDURE
Perform the Test
1) From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601PRO front
panel outlet, and turn on the device.
2) Attach the device's applied parts to the 601PRO applied part terminals if applicable.
3) Press shortcut key 4.The Earth Leakage test appears on the display, and the test begins
immediately:
SOFT KEY 1 toggles the DUT outlet Polarity from Normal to Off to Reverse.
SOFT KEY 2 toggles the DUT outlet from Earth to No Earth.
SOFT KEY 3 toggles the DUT outlet from L2 to No L2.
SOFT KEY 4 toggles the AP to Earth to No AP to Earth.
4) Press the print data key at any time to generate a printout of the latest measurement.
Summary of Contents for DC-68
Page 2: ......
Page 12: ......
Page 18: ......
Page 30: ......
Page 45: ...System Installation 3 15...
Page 54: ......
Page 72: ......
Page 90: ......
Page 108: ...8 2 Field Replaceable Unit 8 1 Explosive View A0 B0 C0 D0 E0 F0 G0...
Page 121: ...Field Replaceable Unit 8 15 8 2 5 Main Unit Assembly E0 E2 E4 E6 E5 E9 E8 E3 E7 E1...
Page 123: ...Field Replaceable Unit 8 17 E13 E22 E10 E17 E16 E14 E18 E19 E23 E20 E21 E15 E11 E12 E24...
Page 134: ......
Page 180: ......
Page 212: ......
Page 254: ......
Page 258: ......
Page 280: ......
Page 281: ...P N 046 011984 00 7 0...