13-4 Troubleshooting of Regular Malfunctions
13.3 Image Problems
13.3.1 Related Modules or Boards
No.
Description
Remarks
1
DC-DC board
Power supply part
2
CW board
CW mode
3
TP64 board
/
5
Engine board
Signal process of the ultrasound system
6
Probe board assembly
/
13.3.2 Key Points Supporting Troubleshooting
No.
Key Points Supporting Troubleshooting
Remarks
1
Image feature, including dark strips and noise
Describe image characteristics in the
following two lists.
2
Image appearance for varied probes connecting to
varied sockets.
/
3
Imaging characteristics for varied probes
connecting to varied/same interfaces.
It may happen because of any relay
failure for probe switching or probe
element switching.
4
Self test for Main Voltage Test
/
5
Self test for ATGC Function Test
/
6
Self test for AFE Noise Test
/
7
Self test for AFE Digital Interface Test
/
8
Self test for CPU Base Board and Engine Board
Interconnection test
/
9
Self test for Engine Board and TR64 Board
Interconnection test
/
10
Self test for TR64 Board and CW Board
Interconnection test
/
Summary of Contents for DC-68
Page 2: ......
Page 12: ......
Page 18: ......
Page 30: ......
Page 45: ...System Installation 3 15...
Page 54: ......
Page 72: ......
Page 90: ......
Page 108: ...8 2 Field Replaceable Unit 8 1 Explosive View A0 B0 C0 D0 E0 F0 G0...
Page 121: ...Field Replaceable Unit 8 15 8 2 5 Main Unit Assembly E0 E2 E4 E6 E5 E9 E8 E3 E7 E1...
Page 123: ...Field Replaceable Unit 8 17 E13 E22 E10 E17 E16 E14 E18 E19 E23 E20 E21 E15 E11 E12 E24...
Page 134: ......
Page 180: ......
Page 212: ......
Page 254: ......
Page 258: ......
Page 280: ......
Page 281: ...P N 046 011984 00 7 0...