9-2 Structure and Assembly/Disassembly
No.
Name
No.
Name
<11>.
Back cover assembly of main unit
<12>
The handle of main unit
<13>.
Display cable cover
<14>
Support arm assembly
<15>.
Display (monitor) assembly
/
9.2 Preparation
9.2.1 Disassembly Tools Required
Name
Type
Material No.
Remarks
Cross-headed screwdriver
107*75
/
M2
Cross-headed screwdriver
107*75
0000-10-10884
M3
、
M4
Open-ended spanner
8#
042-007605-00
Diagonal cutting pliers
N-206S 095-000077-00
/
Anti-electrostatic glove: 1 pair.
/
/
/
9.2.2 Engineers
Required
The disassembly should be performed by professionals from Mindray or the staff who are qualified
for the maintenance after the training.
9.2.3 Disassembly Requirements
Be prepared before disassembling ultrasound device.
1. Stop scanning the patient and capturing images. Shut down the device and cut off AC power
supply. Unplug AC power supply cable.
2. Lock the casters to prevent the device from moving in the disassembly.
3. Get the disassembly tools prepared.
9.3 Assembly/Disassembly
This section describes the disassembling and the assembling of the main modules and hardware
boards. The assembling is the inverse process of the disassembling if not mentioned in particular.
Summary of Contents for DC-68
Page 2: ......
Page 12: ......
Page 18: ......
Page 30: ......
Page 45: ...System Installation 3 15...
Page 54: ......
Page 72: ......
Page 90: ......
Page 108: ...8 2 Field Replaceable Unit 8 1 Explosive View A0 B0 C0 D0 E0 F0 G0...
Page 121: ...Field Replaceable Unit 8 15 8 2 5 Main Unit Assembly E0 E2 E4 E6 E5 E9 E8 E3 E7 E1...
Page 123: ...Field Replaceable Unit 8 17 E13 E22 E10 E17 E16 E14 E18 E19 E23 E20 E21 E15 E11 E12 E24...
Page 134: ......
Page 180: ......
Page 212: ......
Page 254: ......
Page 258: ......
Page 280: ......
Page 281: ...P N 046 011984 00 7 0...