A-6 Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
4- Protective Earth Resistance
2) Attach the 601PRO RED input lead to the device’s Protective Earth terminal or an exposed
metal area.
3) Press shortcut key 3. The Protective Earth Resistance test is displayed.
4) Press SOFT KEY 3 to select a test current (1AMP, 10AMP, or 25AMP). The selected test
current is displayed in the upper right corner of the display.
5) Press START TEST to start the test. The test current is applied while resistance and current
readings are taken. This takes approximately 5 seconds.
6) Press the print data key at any time to generate a printout of the latest measurement(s).
NOTE
When "Over" is displayed for Ohms, this signifies that a valid measurement was not obtained
because either an open connection was detected or that the measurement was not within range.
Readings greater than 9.999 Ohms will be displayed as Over.
Failure
Once it reaches the limitation, stop using equipment. Check the protective earth connection
between
Protective Earth terminal and Protective Earth Metal enclosure and equipotential terminal;
Retest
and inform the Customer Service Engineer for analysis and disposal if still fail.
LIMITS
ALL COUNTRIES R = 0.2
Ω
Maximum
Summary of Contents for DC-68
Page 2: ......
Page 12: ......
Page 18: ......
Page 30: ......
Page 45: ...System Installation 3 15...
Page 54: ......
Page 72: ......
Page 90: ......
Page 108: ...8 2 Field Replaceable Unit 8 1 Explosive View A0 B0 C0 D0 E0 F0 G0...
Page 121: ...Field Replaceable Unit 8 15 8 2 5 Main Unit Assembly E0 E2 E4 E6 E5 E9 E8 E3 E7 E1...
Page 123: ...Field Replaceable Unit 8 17 E13 E22 E10 E17 E16 E14 E18 E19 E23 E20 E21 E15 E11 E12 E24...
Page 134: ......
Page 180: ......
Page 212: ......
Page 254: ......
Page 258: ......
Page 280: ......
Page 281: ...P N 046 011984 00 7 0...