Care and Maintenance 12-1
12
Care and Maintenance
12.1 Overview
The maintenance procedure in this chapter is only for recommendation.
12.1.1 Tools, Measurement Devices and Consumables
Table 12-1 The list for Tools and Measurement Devices
Tools/Measurement
Devices
Qty.
Remarks
Plastic and resin
container
1
Used to contain the physiological saline and two
probes available in the container.
Soft brush
1
The size is similar to that of the brush. The
material of the brush should be soft.
Small plastic bowl
1
Used to contain soap-suds
Safety testing device
1
See Appendix A for details
Table 12-2 Consumables List
12.1.2 Routine Maintenance Items
Table 12-3 The List for Maintenance Items and Maintenance Frequency
No.
Item
Frequency
Method
1.
Dust-proof cover cleaning
1 time/month
See
Chapter 12.2.1
2.
The monitor and the touch screen
cleaning
1 time/month
Ditto
Consumable
Qty.
Remarks
Aluminum foil
About 1 meter long /
Physiological saline
About 1000 mL
Reach the half of the container to submerge
the probe in the saline.
(See Appendix A for details)
(Concentration 0.85% to 0.95%)
Mild soap-suds
About 400 mL
/
Dry soft cloth
5 pieces
/
Summary of Contents for DC-68
Page 2: ......
Page 12: ......
Page 18: ......
Page 30: ......
Page 45: ...System Installation 3 15...
Page 54: ......
Page 72: ......
Page 90: ......
Page 108: ...8 2 Field Replaceable Unit 8 1 Explosive View A0 B0 C0 D0 E0 F0 G0...
Page 121: ...Field Replaceable Unit 8 15 8 2 5 Main Unit Assembly E0 E2 E4 E6 E5 E9 E8 E3 E7 E1...
Page 123: ...Field Replaceable Unit 8 17 E13 E22 E10 E17 E16 E14 E18 E19 E23 E20 E21 E15 E11 E12 E24...
Page 134: ......
Page 180: ......
Page 212: ......
Page 254: ......
Page 258: ......
Page 280: ......
Page 281: ...P N 046 011984 00 7 0...