MI 3290 Earth Analyser
Instrument operation
23
Previous screen view.
Next screen view.
Previous screen result or increase parameter.
Next screen result or decrease parameter.
Edit chart (zoom in or zoom out, and move cursor).
Opens help screen.
Views results of measurement.
Starts test leads compensation in
Ω - Meter (200 mA and 7 mA) measurement.
Starts balancing procedure in Selective (Flex Clamps 1-4) measurement.
Expands control panel / open more options.
Warning!
High voltage is applied to the test terminals. Measurement will not be
started.
Limit [ > 50 Vrms H-E, S-E, ES-E, H-Guard, S-Guard, ES-Guard ].
The measuring range of the instrument is exceeded.
Measurement will not be started or displayed!
High electrical noise was detected during measurement. Results may be
impaired.
Limit [ Noise frequency is close
(±6 %) to the test frequency ].
Measurement is running, consider displayed warnings.
High impedance to earth of test probes.
Influence of the auxiliary electrodes
High impedance of current probe Rc.
Influence of the auxiliary electrodes
High impedance of current probe Rp.
15.9 Influence of the auxiliary electrodes.
Test leads resistance in Ω - Meter (200 mA and 7 mA) measurement is not
compensated.
Limit [ Lead compensation <
5 Ω ].