MI 3290 Earth Analyser
Appendix F - The balancing procedure
158
Show picture
browse for image path
Duration Number in seconds, infinite
no entry
Buzzer mode
Passed or failed measurement is indicated with beeps.
Pass
– double beep after the test
Fail
– long beep after the test
Beep happens right after single test measurement.
Parameters:
State
On
– enables Buzzer mode
Off
– disables Buzzer mode
Operation after end of test
This flow command controls the proceeding of the Auto Sequence
® in regard to the
measurement results.
Parameters:
Operation after end of test
– pass
– fail
– no status
The operation can be individually set for the case the
measurement passed, failed or ended without a
status.
Manual
– The test sequence stops and waits for
appropriate command (Enter key) to
proceed.
Auto
–
The
test
sequence
automatically
proceeds.