MI 3290 Earth Analyser
Tests and Measurements
98
Test parameters for
Ω - Meter (7 mA):
Sound
[On, Off]
Limit (R)
Limit value selection: [
OFF, 1 Ω – 15.0 kΩ]
Ω - Meter (7 mA) measurement procedure:
Select the Ω - Meter (7 mA) measurement function.
Set the test parameters (sound and limit).
Connect the test leads to the instrument.
Compensate the leads (optional).
Connect the test leads to the test object.
Press the Run key to start the measurement.
Wait until the test result is displayed on the screen.
Press the Run key to stop the measurement.
Save results (optional).
Note:
Consider displayed warnings when starting the measurement!
11.5.2.1 Compensation of test leads resistance
This chapter describes how to compensate test leads resistance in both continuity functions
(
Ω - Meter 200 mA and 7 mA). Compensation is required in 2-wire mode to eliminate the
influence of test leads resistance and the internal resistances of the instrument on the
measured resistance. The lead compensation is therefore a very important feature to obtain
correct result. Once compensation has been performed, the compensation icon
appears on the screen.
Circuits for compensating the resistance of test leads
Figure 11.59: Shorted test leads
Compensation of test leads resistance procedure:
Select the Ω - Meter 200 mA or 7 mA function.
Connect test cable to the instrument and short the test leads together see
Press the
icon to compensate leads resistance.
Notes:
The limit value for lead compensation is 5 Ω.
The lead compensation current is 200 mA DC.