MI 3290 Earth Analyser
Tests and Measurements
111
11.9.1 Iron Clamp Meter RMS Measurement
Test can be started from the Iron Clamp Meter RMS measurement window. Before carrying
out a test the following parameters (Measurement Clamp Type and Limit (Ic)) can be edited.
Figure 11.90: Iron Clamp Meter RMS measurement menu
Test parameters for Iron Clamp Meter RMS:
Measurement
Clamp type
Set clamp type: [A1018]
Limit (Ic)
Limit value selection: [OFF, 10 mA
– 9.00 A]
Iron Clamp Meter RMS measurement procedure:
Select the Iron Clamp Meter RMS measurement function.
Set the test parameter (clamp type and limit).
Connect the clamp to the instrument and to the test object.
Press the Run key to start the measurement.
Wait until the test result is displayed on the screen.
Press the Run key to stop the measurement.
Save results (optional).
Figure 11.91: Example of Iron Clamp Meter RMS
measurement result
Note:
Consider displayed warnings when starting the measurement!