MI 3152(H)
Eurotest
XC (2,5 kV)
General description
8
The instrument comes supplied with rechargeable Ni-MH battery cells. The cells
should only be replaced with the same type as defined on the battery
compartment label or as described in this manual. Do not use standard alkaline
battery cells while the power supply adapter is connected, otherwise they may
explode!
Hazardous voltages exist inside the instrument. Disconnect all test leads, remove
the power supply cable and switch off the instrument before removing battery
compartment cover.
Do not connect any voltage source on C1/C2 inputs. It is intended only for
connection of current clamps. Maximal input voltage is 3 V!
1.1.2 Markings on the instrument
Read the Instruction manual with special care to safety operation«. The
symbol requires an action!
Mark on your equipment certifies that it meets requirements of all
subjected EU regulations.
This equipment should be recycled as electronic waste.
1.1.3 Warnings related to safety of batteries
When connected to an installation, the instruments battery compartment can contain
hazardous voltage inside! When replacing battery cells or before opening the
battery/fuse compartment cover, disconnect any measuring accessory connected to the
instrument and turn off the instrument,
Ensure that the battery cells are inserted correctly otherwise the instrument will not
operate and the batteries could be discharged.
Do not recharge alkaline battery cells!
Use only power supply adapter delivered from the manufacturer or distributor of the test
equipment!
1.1.4 Warnings related to safety of measurement functions
Insulation resistance
Insulation resistance measurement should only be performed on de-energized objects!
Do not touch the test object during the measurement or before it is fully discharged! Risk
of electric shock!
Continuity functions
Continuity measurements should only be performed on de-energized objects!