MI 3152(H)
Eurotest
XC (2,5 kV)
Technical specifications
177
12.2
Diagnostic test (MI 3152H only)
Uiso: 500V, 1000 V, 2500 V
DAR
– Dielectric absorption ratio
Measuring range
Resolution
Accuracy
0.01 ... 9.99
0.01
(5 % of r 2 digits)
10.0 ... 100.0
0.1
(5 % of reading)
PI
– Polarization index
Measuring range
Resolution
Accuracy
0.01 ... 9.99
0.01
(5 % of r 2 digits)
10.0 ... 100.0
0.1
(5 % of reading)
For
Riso
,
R60,
and
Um
sub-results technical specifications defined in chapter
apply.
12.3
R low – Resistance of earth connection and
equipotential bonding
Measuring range according to EN 61557 is 0.16
... 1999
.
R
– Resistance
Measuring range (
)
Resolution (
)
Accuracy
0.00 ... 19.99
0.01
(3 % of r 3 digits)
20.0 ... 199.9
0.1
(5 % of reading)
200 ... 1999
1
R+, R
– Resistance
Measuring range (
)
Resolution (
)
Accuracy
0.0 ... 199.9
0.1
(5 % of r 5 digits)
200 ... 1999
1
Open-circuit voltage ................................ 6.5 VDC ... 18 VDC
Measuring current ................................... min. 200 mA into load resistance of 2
Test lead compensation .......................... up to 5
The number of possible tests .................. > 1400, with a fully charged battery
Automatic polarity reversal of the test voltage.
12.4
Continuity – Continuous resistance measurement with
low current
R
– Continuity resistance
Measuring range (
)
Resolution (
)
Accuracy
0.0 ... 19.9
0.1
(5 % of r 10 digits)
20 ... 1999
1
Open-circuit voltage ................................ 6.5 VDC ... 18 VDC
Short-circuit current ................................ max. 8.5 mA