MI 3152(H)
Eurotest
XC (2,5 kV)
Technical specifications
190
12.25
Illumination
Illumination (A 1172)
Specified accuracy is valid for complete operating range.
Measuring range (lux)
Resolution (lux)
Accuracy
0.01 ... 19.99
0.01
(5 % of r 2 digits)
20.0 ... 199.9
0.1
(5 % of reading)
200 ... 1999
1
2.00 ... 19.99 k
10
Measurement principle ........................... silicon photodiode with V(
) filter
Spectral response error .......................... < 3.8 % according to CIE curve
Cosine error ............................................ < 2.5 % up to an incident angle of
85
O
Overall accuracy ..................................... matched to DIN 5032 class B standard
Illumination (A 1173)
Specified accuracy is valid for complete operating range.
Measuring range (lux)
Resolution (lux)
Accuracy
0.01 ... 19.99
0.01
(10 % of r 3 digits)
20.0 ... 199.9
0.1
(10 % of reading)
200 ... 1999
1
2.00 ... 19.99 k
10
Measurement principle ........................... silicon photodiode
Cosine error ............................................ < 2.5 % up to an incident angle of
85
O
Overall accuracy ..................................... matched to DIN 5032 class C standard
12.26
Auto Sequences®